| Cat.No. | Model | Measurement wavelength (λ) | Measurement range | Temperature control range | Speed control range |
|---|---|---|---|---|---|
| R11-650-020 | RTS-8 | 850 ± 15 nm | 0-100 OD600 | +15 ℃ below ambient ... +60 ℃ | 150-2700 rpm |
| Cat .No. | R11-650-020 |
|---|---|
| Model | RTS-8 |
| Light source | Laser |
| Measurement wavelength (λ) | 850 ± 15 nm |
| Measurement range | 0-100 OD600 |
| E.coli factory calibration measurement range | 0-50 OD600 |
| S.cerevisiae factory calibration measurement range | 0-75 OD600 |
| Achievable user calibration measurement error (range 0.1-6 OD600) | ±0.3 |
| Achievable user calibration measurement error (range 6-50 OD600) | ≤ 5% |
| Achievable user calibration measurement error (range 50-75 OD600) | ≤ 10% |
| Measurement periodicity per hour | 1-60 |
| Temperature setting range | +15℃... +60℃ |
| Temperature control range | +15 ℃ below ambient ... +60 ℃ |
| Temperature stability | ±0.3 ℃ |
| Sample temperature accuracy (20℃-37℃) | ±1 ℃ |
| Tube sockets | 8 |
| Sample working volume range | 3-50 ml |
| Speed control range | 150-2700 rpm |
| Reverse spin time setting range 150-250 rpm | 0 s |
| Reverse spin time setting range 250-300 rpm | 2-60 s |
| Reverse spin time setting range 300-2700 rpm | 0-60 s |
| Display | LCD |
| Minimum PC requirements | Intel/AMD Processor, 1 GB RAM, Windows Vista/7/8/8.1/10, USB 2.0 port |
| Dimensions (W×D×H) | 350 ×690× 300 mm |
| Weight | 20 kg |
| Nominal operating voltage | AC 230 V, 50 Hz |
| Power consumption | 3.15 A / 500 W |